Bowie-Dick Test 134°C/3.5 min – Helix PCD Kit with 100 Indicator Strips for Steam Sterilizer Daily Control, ISO 11140-1 Class 2, SPS MEDICAL
- Bowie-Dick Test 134°C/3.5 min (100 pcs) – professional daily autoclave control. Detects sterilization errors. For medical, dental, beauty, and tattoo use. SPS MEDICAL. Discover full details – click for more!
Subscription products in your shopping cart
Set one delivery frequency for all subscription products in your cart every:
Subscription products in your shopping cart
Bowie-Dick Test 134°C/3.5 min. (100 pcs) – Professional Set for Autoclave Quality Control
Professional Bowie-Dick Test – Assurance of Safety and Sterilization Effectiveness
Bowie-Dick Test 134°C/3.5 min. is a comprehensive set for daily quality control of vacuum autoclaves. The kit includes a specialized test device (PCD) and 100 class 2 indicator strips designed for sterilization cycles at 134°C for 3.5 minutes. This product is dedicated for professional use in medical, dental, podiatry clinics, beauty salons, tattoo studios, and other industries requiring the highest standards of tool sterilization.
With the Bowie-Dick test, you can quickly and effectively detect any irregularities in autoclave operation, such as too low temperature, insufficient sterilization time, inadequate vacuum, poor air removal, insufficient steam penetration, or the presence of condensate or inert gases. The test also enables early detection of leaks and system faults within the autoclave.
Each indicator strip features a chemical indicator that changes color from blue to pink after a correct sterilization process, providing clear and unambiguous result interpretation. The set allows you to conduct controls in accordance with applicable standards and guidelines, and also facilitates documentation of sterilization processes.
All products offered in the autoclaves.shop store are specially selected for professional use in various fields – from medicine and dentistry to podiatry, beauty, and tattoo industries. We guarantee the highest quality, full compliance with standards (including ISO 11140-1 class 2), complete documentation, and compatibility with a wide range of devices available on the market. Our store has been operating continuously since 2005, being the first on the autoclave market in Poland, now serving dozens of countries worldwide.
Main Product Features:
- Set includes: 1 Helix test device (polypropylene tube with 1.5 m Teflon tube), 100 indicator strips, cotton bag, and instructions with an illustration of the indicator color change
- Self-adhesive indicator strips, width 6 mm, length 38 mm (total test with description 7.6 cm)
- Operating range: 134°C for 3.5 minutes (dry saturated steam)
- Clear color change of the indicator from blue to pink after a proper cycle
- Dedicated for all types of steam sterilizers
- Full compliance with technical standards and requirements for sterilization of medical, cosmetic, podiatry, and tattoo tools
- Cardboard packaging: 178 x 125 x 50 mm
- Each package features a label with product description, contents, reference number, LOT number, production date, and expiration date
Instructions for Use and Result Interpretation:
The Bowie-Dick test should be placed in the autoclave before starting work each day, after servicing, relocating the device, or after a prolonged break in use. After the cycle is complete (3.5 minutes at 134°C), the indicator strip should change color from blue to pink. Lack of full color change or uneven coloration indicates issues with autoclave performance.
Product Parameters
- Product code: 101.201.0100
- Manufacturer: SPS MEDICAL
- Standard compliance: ISO 11140-1 class 2
- Set contents: PCD device (Helix) + 100 indicator strips + cotton bag + instructions
- Package dimensions: 178 x 125 x 50 mm
- Application area: All types of steam sterilizers
Why Choose Products from autoclaves.shop?
- Products specially selected for professional use in medicine, cosmetology, podiatry, and the beauty industry
- Top quality, compliance with standards, and complete documentation
- Compatibility of spare parts and consumables with devices is clearly specified in the Compatibility section below the product description, making it easy to make the right purchase
- The store has been operating continuously since 2005 – the first on the autoclave market in Poland, now serving dozens of countries worldwide