HLP 10sp water demineralizer
For lab, medical and dental...
HLP series demineralizers are ideal tap water-fed devices capable of meeting all the requirements of any modern laboratory for the need of ultrapure water suitable for analytical and instrumental purposes.
Demineralizer HLP 10sp. General information
- Supply: tap water
- Capacity: 10 l/h
- Feed rate of treated water: 1-2 l/min
- Conductivity: < 0.06 μS/cm
- Bacteria: < 1cfu/ml
- Particles: > 0.2µm < 1/ml
Standard: water treated in the device meets the requirements of PN-EN ISO 3696: 1999 for water of the first, second and third degree of purity, corresponds microbiologically and physicochemically to the requirements of the FP for treated production water.
Application: the water obtained can be used for instrumental analysis of AAS, ICP/MS, IC, HPLC, GC, bacterial culture, biochemical analysis.
The device is equipped with a 0.45/0.2μm cascade microfiltration capsule and a feed pressure booster pump.
Technical parameters
- The device works under the pressure of tap water.
- Water purification stages:
- 5µm sediment prefilter,
- sediment-carbon filtration and softening (integrated module A),
- reverse osmosis,
- double demineralization on a mixed ion exchange bed,
- 0.45/0.2μm cascade microfiltration capsule.
- System performance ok. 10 - 12 dm3/h (daily approx. 240-280 dm3).
- Conductivity of demineralized water < 0.06 µS/cm.
- Automatic and unattended operation of the device.
- The device is equipped with a pump that raises the supply pressure,
- The system is equipped with a pressure vessel with a capacity of. 10 dm3.
- Possibility to replace the tank with a larger capacity tank.
- Two independent water intake points of Class I and Class II:
- Mobile water intake point - second class of purity according to PN-EN ISO 3696: 1999; provided with demi water tap with a range of min. 2 m and a pressure vessel of. 10 dm3,
- Mobile water intake point of the first purity class according to PN-EN ISO 3696:1999 (with microfiltration capsule 0.45/0.2μm).
- Installation of an additional point of intake of general laboratory water - third class of cleanliness according to PN-EN ISO 3696: 1999 (for washing glassware, supplying laboratory dishwasher, autoclave, etc.).). (optional)
- Automatic stoppage of system operation when the tank is full.
- Self-service capability (easy replacement of replaceable modules).
- Maximum working pressure: 1 MPa.
- The system is designed for cold water supply: 5-40°C.
- Power supply: 230V/50Hz.
- Possibility of self-assembly of the device.
Functions that monitor system operation:
- The device is equipped with 24V automation with a microprocessor-based control and measurement system having:
- color LCD display,
- a conductivity meter that measures the conductivity and temperature of purified water in units of µS/cm or MOhm,
- temperature compensation,
- a clock that displays the date and time,
- information about the current operating status of the system,
- software
- feedwater pressure gauge.
- graphic and audible alarm signaling:
- an alarm indicating the replacement of module A,
- an alarm indicating the replacement of the ion exchange module,
- alarm to inform about replacement of microfiltration capsule,
- preview of service appointments,
- modification of the contact service telephone number by the User,
- built-in RS 232 interface for communication with a computer,
- possibility of individual adjustment of service frequency and alarm levels directly from the device keyboard.
Dimensions (width. x deep. x height.): 230x360x520 mm
Feedwater parameters:
- Conductivity: < 1200 µS/cm
- Pressure: > 3.5 bar
- Temperature: 4-40 ºC
- Hardness: < 250 mg CaCO3/dm3
- Iron: < 0.2 mg/dm3
On-site connections:
- cold tap water connection ½" or ¾"
- 230V socket
- drain
Consumables: